Extraction of Reliable Information from Time-Domain Pressure and Flow Signals Measured By Means of Forced Oscillation Techniques

Показать сокращенную информацию

dc.contributor Казанский федеральный университет
dc.contributor.author Nigmatullin Ravil Rashidovich
dc.contributor.author Osokin Sergey Igorevich
dc.contributor.author Ionescu Clara
dc.contributor.author Baleanu Dumitru
dc.date.accessioned 2024-12-04T07:42:33Z
dc.date.available 2024-12-04T07:42:33Z
dc.date.issued 2014
dc.identifier.citation Extraction of Reliable Information from Time-Domain Pressure and Flow Signals Measured By Means of Forced Oscillation Techniques / Raoul R. Nigmatullin, Serghei I. Osokin, Clara Ionescu, Dumitru Baleanu // REVISTA DE CHIMIE. - 2014. - Vol.65. - No. 3. - P. 299-305.
dc.identifier.uri https://dspace.kpfu.ru/xmlui/handle/net/184983
dc.description.abstract This paper aims to give a proof-of-concept for the possible application of the forced oscillation lung function test to assess the viscoelastic properties of the airways and tissue. In particular, a novel signal processing algorithm is employed on non-stationary, noisy, (relatively) short time series of respiratory pressure and flow signals. This novel technique is employed to filter the useful information from the signals acquired under two measurement conditions: pseudo-functional residual capacity (PFRC) and pseudo-total lung capacity (PTLC). The PFRC is the measurement performed at lowest lung volume with maximum deflation, and the PTLC is measurement performed at the maximum lung volume under maximum inflation. The results suggest that the proposed technique is able to extract information on the viscoelastic properties of the lung tissue at a macroscopic level. The conclusion of this preliminary study is that the proposed combination of signal processing method and lung function test is suited to be employed on a large database in order to deliver reference values and perform further statistical analysis.
dc.language.iso en
dc.relation.ispartofseries REVISTA DE CHIMIE
dc.rights открытый доступ
dc.subject respiratory mechanics
dc.subject detrending procedure
dc.subject filtering
dc.subject lung function
dc.subject oscillatory mechanics
dc.subject statistical signal processing
dc.subject.other Математика
dc.subject.other Физика
dc.title Extraction of Reliable Information from Time-Domain Pressure and Flow Signals Measured By Means of Forced Oscillation Techniques
dc.type Article
dc.contributor.org Институт информационных технологий и интеллектуальных систем
dc.description.pages 299-305