Показать сокращенную информацию
dc.contributor.author | Alakshin E. | |
dc.contributor.author | Gazizulin R. | |
dc.contributor.author | Klochkov A. | |
dc.contributor.author | Kuzmin V. | |
dc.contributor.author | Sabitova A. | |
dc.contributor.author | Safin T. | |
dc.contributor.author | Tagirov M. | |
dc.date.accessioned | 2018-09-18T20:35:41Z | |
dc.date.available | 2018-09-18T20:35:41Z | |
dc.date.issued | 2013 | |
dc.identifier.uri | https://dspace.kpfu.ru/xmlui/handle/net/141496 | |
dc.description.abstract | The home-built pulse nuclear magnetic resonance spectrometer for 3He research is described. The temperature range is 1.5 - 4.2 K, the frequency range is 3 - 20 MHz, the dead time is 10 μs at 10 MHz. The spectrometer software is based on programming language LabVIEW. The advantages of digital technique are described; particularly the implementation of digital quadrature detector. © Kazan Federal University (KFU). | |
dc.subject | 3He | |
dc.subject | Digital detection | |
dc.subject | Low temperature | |
dc.subject | Nuclear magnetic resonance | |
dc.subject | Spectrometer | |
dc.title | The home-built pulse nuclear magnetic resonance spectrometer with digital quadrature detection for 3He research at low temperatures | |
dc.type | Article | |
dc.relation.ispartofseries-issue | 1 | |
dc.relation.ispartofseries-volume | 15 | |
dc.collection | Публикации сотрудников КФУ | |
dc.relation.startpage | 1 | |
dc.source.id | SCOPUS-2013-15-1-SID84879025313 |