dc.description.abstract |
© 2015, Springer Science+Business Media New York. A voltammetric method for determining acyclovir using a chemically modified electrode with catalytic properties was developed. The catalytic properties for acyclovir oxidation of glassy-carbon electrodes modified by inorganic films of Ru(III) hexachloroplatinate or hexacyanocobaltate were compared. The catalytic effect was manifested as a decreased potential and multiply increased oxidation current at the proposed film electrodes. The greatest catalytic effect was observed using an electrode with a film of Ru(III) hexachloroplatinate. The catalytic current depended linearly on the analyte concentration in the range from 0.5 μM to 5 mM. The proposed method was used to determine acyclovir in drugs. |
|
dc.relation.ispartofseries |
Pharmaceutical Chemistry Journal |
|
dc.subject |
acyclovir |
|
dc.subject |
inorganic films |
|
dc.subject |
modified electrode |
|
dc.subject |
ruthenium hexachloroplatinate and hexacyanocobaltate |
|
dc.subject |
voltammetric determination |
|
dc.title |
Voltammetric Determination of Acyclovir in Drugs Using an Electrode Modified by Ruthenium Hexachloroplatinate or Hexacyanocobaltate Film |
|
dc.type |
Article |
|
dc.relation.ispartofseries-issue |
11 |
|
dc.relation.ispartofseries-volume |
48 |
|
dc.collection |
Публикации сотрудников КФУ |
|
dc.relation.startpage |
747 |
|
dc.source.id |
SCOPUS0091150X-2015-48-11-SID84942089788 |
|