dc.description.abstract | An alternative method is developed for recording atomic absorption with spatial resolution on the basis of a position-sensitive photodetector. The main requirements on the optical system, the source, the radiation detector, and the algorithms for processing and interpreting the results are formulated. A description is given of how the proposed method is put into practice on the basis of a commercial atomic-absorption spectrophotometer. The advantages of the new recording system are shown, using the atomization of biological specimens of complex composition as an example. | |