dc.contributor.author |
Kazakov B. |
|
dc.contributor.author |
Mikheev A. |
|
dc.contributor.author |
Motygullin I. |
|
dc.date.accessioned |
2018-09-17T20:16:42Z |
|
dc.date.available |
2018-09-17T20:16:42Z |
|
dc.date.issued |
2005 |
|
dc.identifier.issn |
0020-4412 |
|
dc.identifier.uri |
https://dspace.kpfu.ru/xmlui/handle/net/133275 |
|
dc.description.abstract |
It was shown in experiments that the application of a differentiating RC circuit with a time constant comparable to the rise time of the single-electron pulse in the amplifying section of a photon counting system increases the linearity range of the light characteristic. In this case, the maximum recorded intensity of the light flux increases approximately by a factor of 10, as compared with the maximum intensity measured without the RC circuit.© 2005 Pleiades Publishing, Inc. |
|
dc.relation.ispartofseries |
Instruments and Experimental Techniques |
|
dc.title |
A differentiating RC circuit in the amplifying section of a photon counting system |
|
dc.type |
Article |
|
dc.relation.ispartofseries-issue |
3 |
|
dc.relation.ispartofseries-volume |
48 |
|
dc.collection |
Публикации сотрудников КФУ |
|
dc.relation.startpage |
308 |
|
dc.source.id |
SCOPUS00204412-2005-48-3-SID22344445813 |
|