Показать сокращенную информацию
dc.contributor.author | Kashirin N. | |
dc.contributor.author | Skirda V. | |
dc.contributor.author | Idiyatullin D. | |
dc.contributor.author | Sevryugin V. | |
dc.date.accessioned | 2018-09-17T20:16:27Z | |
dc.date.available | 2018-09-17T20:16:27Z | |
dc.date.issued | 1998 | |
dc.identifier.issn | 0020-4412 | |
dc.identifier.uri | https://dspace.kpfu.ru/xmlui/handle/net/133269 | |
dc.description.abstract | An active system of temperature regulation and temperature gradient in a sample on the example NMR sensor is described in this work. By supporting the given average temperature, the system of temperature-stabilization allows one to set and supervise the size and direction of the temperature gradient along a sample. The temperature of a sample is maintained with on accuracy of about ±0.1°C in a temperature interval of-150 to +200°C with a longitudinal temperature gradient no more than ±0.2°C. The value of the temperature gradient is established in the range of 0° to 10°C/cm in both directions. © 1998 MAHK Hayka/Interperiodica Publishing. | |
dc.relation.ispartofseries | Instruments and Experimental Techniques | |
dc.title | System of active regulation of the temperature and temperature gradient in nuclear magnetic resonance sensors | |
dc.type | Article | |
dc.relation.ispartofseries-issue | 3 | |
dc.relation.ispartofseries-volume | 41 | |
dc.collection | Публикации сотрудников КФУ | |
dc.relation.startpage | 443 | |
dc.source.id | SCOPUS00204412-1998-41-3-SID27544480488 |