dc.description.abstract |
A circuit for setting and regulating specimen temperature in NMR spectrometers based on a full-wave rectifier and a thermocouple sensor is described. Temperature fluctuations do not exceed 0. 2 degree K in the range of 473-123 degree K. |
|